Medical devices used to be islands. A pacemaker talked to its programmer in the cardiologist's office over a wand placed against the patient's chest. An infusion pump in a hospital room had a serial port for service technicians. Insulin pumps were standalone. None of these devices touched a network and most did not even include encryption.
That world is gone. Modern pacemakers transmit telemetry over Bluetooth Low Energy and cellular. Infusion pumps connect to hospital information systems over Wi-Fi to receive medication orders and report alerts. Continuous glucose monitors stream data to smartphones. Every connected medical device is now a target, and every target needs cryptographic protection that will hold up over the device's full operational lifetime.
That lifetime is the problem. A pacemaker is implanted for ten to fifteen years. An infusion pump is in service for seven to ten years. A CT scanner runs for over a decade between major upgrades. Whatever cryptography ships in these devices today must still be secure when they retire. With NIST publishing post-quantum standards in 2024 and quantum computers expected to threaten classical cryptography in the coming decades, the math is clear. Medical devices need post-quantum cryptography now, not when the threat materializes.
This article walks through the regulatory framework, the engineering constraints, and the practical path forward.
FDA premarket cybersecurity expectations
The Food and Drug Administration treats cybersecurity as a quality and safety issue under 21 CFR Part 820 and related medical device regulations. The 2023 final guidance Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions sets clear expectations for any device submitted to the FDA.
Manufacturers must include a cybersecurity risk assessment in their premarket submission. They must demonstrate a secure product development framework, an SBOM listing all software components, and a plan for postmarket vulnerability management. The 2022 Omnibus Bill amended Section 524B of the Federal Food, Drug, and Cosmetic Act to require these elements explicitly.
Cryptographic agility is a recurring theme in FDA guidance. Devices should be designed so that algorithms can be updated as standards evolve. The agency does not currently mandate post-quantum cryptography, but it has signaled that hybrid implementations using both classical and post-quantum algorithms are an acceptable path forward. Manufacturers who design for agility now will have an easier time meeting future requirements.
The Medical Device Coordination Group in the European Union has issued similar guidance under the Medical Device Regulation. The substantive requirements overlap heavily with FDA expectations, although the documentation formats differ.
IEC 62304 software lifecycle
IEC 62304 is the international standard for medical device software lifecycle. It defines safety classification, risk management, and verification requirements that must be met before software ships in a medical device. Class A is non-injury risk, Class B is non-serious injury risk, and Class C is serious injury or death risk. A pacemaker is Class C. An electronic prescription terminal is typically Class A.
For Class C devices, every software unit must be specified, designed, implemented, integrated, and verified with documented evidence. Cryptographic libraries are no exception. If you embed a post-quantum implementation in a Class C device, you must demonstrate that the implementation matches the specification, that it has been tested against known answer tests, and that any updates follow a documented change management process.
This is where NIST FIPS 203, 204, and 205 help. Because these standards include detailed specifications and test vectors, manufacturers can build a verifiable supply chain. Reference implementations are available from NIST and from established cryptographic libraries. Pairing ML-KEM with X25519 in a hybrid construction, as QNSQY does, gives device manufacturers a path to demonstrate two independent security properties.
NIST FIPS guide covers the standards in more depth, including which assertions a manufacturer can rely on for compliance documentation.
Compute and memory constraints
A pacemaker is a tiny battery-powered computer. Its processor runs at tens of megahertz, not gigahertz. Its memory is measured in tens to hundreds of kilobytes. Its battery has to last a decade. Every cryptographic operation costs energy, and every joule of energy translates to fewer years of life for the device.
ML-KEM-512, the smallest variant, has public keys around 800 bytes and ciphertexts around 770 bytes. Encapsulation runs in low milliseconds on a modern microcontroller. ML-DSA-44 has signatures around 2400 bytes and runs in tens of milliseconds. These numbers are tractable for a device that already has a Bluetooth radio drawing more current than the cryptographic processor.
The problem is the larger parameter sets. ML-KEM-1024 and ML-DSA-87 are slower and use more memory. SLH-DSA, with its hash-tree structure, has signatures up to 49 kilobytes. For a pacemaker, these are out of reach. Manufacturers will end up using the smaller variants and accepting that they will need to migrate again if cryptanalytic improvements weaken those parameter sets.
Energy budget calculations should include both the algorithm execution cost and the radio cost of transmitting larger keys and signatures. A 3-kilobyte signature transmitted over Bluetooth Low Energy can dominate the energy budget of a small device. Battery-powered cardiac devices in particular are sensitive to this and tend to favor parameter sets that minimize wire size.
Infusion pumps and the hospital network
Infusion pumps live in hospital networks where they receive medication orders, alarm states, and library updates. The Manufacturer Disclosure Statement for Medical Device Security, abbreviated MDS2, is a standardized form that hospitals use to evaluate device security posture before purchase. The MDS2 covers encryption, authentication, audit logging, and patch management.
For infusion pumps, the cryptographic load is moderate. Network traffic uses TLS, increasingly over a hybrid post-quantum cipher suite. Drug library updates are signed, often with a chain that ends at a manufacturer root certificate. Audit logs are signed before transmission to a central monitoring system. Each of these flows can use post-quantum algorithms today.
The IEC 60601 series of standards covers electrical safety and essential performance for medical electrical equipment. IEC 60601-1-12 specifically addresses requirements for medical equipment used in hospital environments. Cryptographic operations must not interfere with essential performance, which means a long signing operation that delays a medication delivery alarm is unacceptable. Manufacturers run timing analyses to confirm that crypto fits in the worst-case timing budget for safety-critical operations.
Diagnostic imaging and DICOM
Diagnostic imaging devices like MRI, CT, and ultrasound generate enormous data volumes. Images are transferred using the DICOM standard, which has its own security extensions for transport and at-rest encryption. DICOM Supplement 142 covers digital signatures for clinical documents, and DICOM Supplement 184 addresses key management.
Migrating DICOM-aware infrastructure to post-quantum cryptography means upgrading PACS systems, modality worklist servers, and viewer applications. Healthcare organizations are starting to ask vendors for cryptographic agility commitments, with hybrid post-quantum support targeted for new product lines. NIST FIPS 203 and FIPS 204 are the algorithms most often named in these requirements.
QNSQY can be used as a file-level encryption tool for DICOM archives that need to leave the trusted DICOM transport. For example, image studies sent to external specialists or research collaborators can be wrapped in QNSQY-encrypted containers, providing post-quantum confidentiality independent of the recipient's DICOM software.
Implantable devices and the long lifetime problem
Implantable medical devices including pacemakers, defibrillators, and neurostimulators have lifetimes of seven to fifteen years. Once implanted, the cryptographic algorithms cannot be replaced without explant surgery, which carries its own risks. This makes algorithm choice unusually consequential.
The defensive answer is hybrid cryptography. By combining a classical algorithm like X25519 or Ed25519 with a post-quantum algorithm like ML-KEM or ML-DSA, the device gets two independent security properties. If the post-quantum algorithm is later found to have a flaw, the classical algorithm still provides security against non-quantum attackers. If a quantum computer eventually breaks the classical algorithm, the post-quantum component still holds.
This is exactly the design QNSQY uses for its file encryption format. The same logic applies to implantable device telemetry. ML-KEM-768 plus X25519 is a defensible default for the next decade, with parameter set increases possible if cryptanalysis improves.
Hybrid encryption explains the construction and the threat model in more detail.
Cybersecurity bill of materials
The FDA's premarket guidance requires a software bill of materials, SBOM, that lists every third-party component including cryptographic libraries. This must include version information and known vulnerabilities. For post-quantum cryptography, the SBOM should record which NIST FIPS revision the implementation targets, which test vectors were used for validation, and which constant-time guarantees the implementation provides.
The Cybersecurity and Infrastructure Security Agency, CISA, has published SBOM guidance that aligns with the FDA's expectations. The format is typically SPDX or CycloneDX, both of which can encode cryptographic component metadata.
Healthcare organizations are increasingly requesting SBOMs from vendors before purchase. For post-quantum readiness assessments, the SBOM is the starting point. A device that already includes a hybrid post-quantum cipher in its TLS stack and a hybrid signature scheme in its update mechanism is ahead of devices that still rely solely on RSA and ECDSA.
Postmarket vulnerability management
A medical device cannot be patched the same way a desktop computer can. Patching might require a service visit, an explant, or at minimum a controlled rollout supervised by clinical engineering. The FDA expects manufacturers to have a documented coordinated vulnerability disclosure process and a plan for security updates throughout the device's supported lifetime.
For cryptographic vulnerabilities specifically, manufacturers should monitor NIST announcements, IETF working group output, and academic publications. The Common Vulnerabilities and Exposures database, CVE, includes cryptographic implementation flaws. Side-channel attacks against post-quantum implementations are an active research area, and new defenses appear regularly.
The hybrid construction provides defense-in-depth against unknown future weaknesses. If a side-channel attack is published against a specific ML-KEM implementation, the X25519 component continues to provide security while the manufacturer prepares a patch. Without hybrid, a single broken algorithm leaves the device exposed.
Regulatory submissions and FDA review
When submitting a medical device for FDA premarket review, cryptographic choices appear in several sections. The cybersecurity risk assessment lists the threats the cryptography defends against. The architecture documentation describes where keys are stored and how cryptographic operations are protected from tampering. The verification documentation includes evidence that algorithms have been tested against authoritative test vectors.
The FDA reviews these submissions with input from cybersecurity reviewers in the Center for Devices and Radiological Health. The agency does not certify specific algorithms, but it does evaluate whether the chosen algorithms are appropriate for the device's risk profile. Hybrid post-quantum cryptography aligned with NIST FIPS 203, 204, and 205 is increasingly the expected baseline for new high-risk submissions.
Frequently asked questions
Are post-quantum algorithms FDA-approved for medical devices? The FDA does not approve specific cryptographic algorithms. It expects manufacturers to use algorithms appropriate for the device's risk profile. NIST FIPS 203, 204, and 205 are increasingly cited as appropriate for medical devices.
Does IEC 62304 require post-quantum cryptography? No. IEC 62304 is a software lifecycle standard, not a cryptographic standard. However, the lifecycle requirements apply to cryptographic components, so any post-quantum library used must be specified, verified, and maintained according to the standard.
Can existing pacemakers be updated to post-quantum cryptography? Generally no. Most implantable devices have firmware update mechanisms but do not have the compute headroom to add new cryptographic algorithms. Updates require a hardware refresh, which means waiting for the next generation.
What is the cryptographic load on a typical infusion pump? Modest. TLS sessions for hospital network connectivity, signature verification for drug library updates, and audit log signing. All of these can use post-quantum algorithms today with minimal performance impact.
Should medical device manufacturers use hybrid post-quantum or pure post-quantum? Hybrid is the safer choice for the long lifetimes of medical devices. Combining classical and post-quantum algorithms provides defense-in-depth against unknown future weaknesses in either family.
Sources
- FDA Cybersecurity in Medical Devices Guidance, https://www.fda.gov/medical-devices/digital-health-center-excellence/cybersecurity
- IEC 62304:2006+AMD1:2015 Medical Device Software Lifecycle Processes, https://www.iec.ch/dyn/www/f?p=103:38:::::FSP_ORG_ID,FSP_APEX_PAGE,FSP_PROJECT_ID:1494,23,103609
- NIST FIPS 203, Module-Lattice-Based Key-Encapsulation Mechanism Standard, https://csrc.nist.gov/pubs/fips/203/final
- NIST FIPS 204, Module-Lattice-Based Digital Signature Standard, https://csrc.nist.gov/pubs/fips/204/final
- NIST FIPS 205, Stateless Hash-Based Digital Signature Standard, https://csrc.nist.gov/pubs/fips/205/final
- IEC 60601-1-12 Medical Electrical Equipment Requirements, https://www.iec.ch/
- Section 524B Federal Food Drug and Cosmetic Act, https://www.fda.gov/medical-devices/digital-health-center-excellence/cybersecurity
Related Articles
- What is post-quantum cryptography
- Hybrid encryption
- NIST FIPS guide
- ML-KEM explained
- Harvest now decrypt later
- 6G Research and PQC: ITU-T and 3GPP Direction
- PQC on Industrial PLCs: IEC 62443 Practicality
- ACH Transfers and Post-Quantum Cryptography
- PQC on Automotive ECUs: AUTOSAR and ISO 21434
Protect Your Data Before Q-Day Arrives
QNSQY's NIST-standardized post-quantum encryption protects files against both current and quantum-era threats.